General Parameters | Item | Specifications |
Power Supply | AC220V, 50/60Hz(Standard) 1320VA; AC110V, 60Hz(Optional) |
Size(L*W*H) | 950*645*(1250~1450)mm |
Maximum Heater Output Power(220VAC) | Air Chamber Heater: 330W Water Tank Heater: 155W |
Auxiliary Network Power Output | AC220V/50Hz, both maximum permissible currents are 220V 1.5A |
Heater Indication | 0~100%, 10% increasing amount |
Temperature Control Mode | Bed temperature mode; The skin temperature mode |
Bed Temperature Control Temperature Range | 25.0~37.0℃; 37.1~39.0℃(Span mode) |
Skin Temperature Control Range | 34.0~37.0℃; 37.1~38.0℃(Span mode) |
Skin Temperature Sensor Display Range | 5.0~65.0℃ |
Heating Time | ≤50min |
Temperature Variability | ≤0.5℃ |
Temperature Uniformity(Mattress in horizontal position) | ≤0.8℃ |
Temperature Uniformity(Mattress in tilted position) | ≤1.0℃ |
Difference Between Average Incubator Temperature and Control Temperature | ≤1.5℃ |
Skin Temperature Sensor Accuracy | ±0.3℃ |
Humidity Control | The running time of the tank after it is filled with distilled water | Humidity below 70%RH, at least 24h |
Water Tank Capacity | 1500ml |
Humidity Display Range | 0%RH~99%RH |
Humidity Control Range | CAH%RH~90%RH(1% per adjustment) |
Humidity Control Accuracy | ±10% |
Oxygen Concentration Control | Oxygen Concentration Display Range | 0%~99% |
Oxygen Concentration Display Resolution | 1% |
Oxygen Concentration Control Accuracy | ±5% oxygen volume concentration |
Oxygen Concentration Setting Range | 21%~60%(1% per adjustment) |
Oxygen Sensor Service Life | The maximum use time is 10000h at 100% oxygen concentration |
Transport Storage | Temperature | -20℃~+55℃ |
Atmospheric Pressure Range | 500hPa~1060hPa |
Relative Humidity | ≤93%RH |
Packing Size/Gross Weight | 1150*800*1600mm 190kg |
The Work Environment | Relative Humidity | 30%RH~75%RH |
The Environment Temperature | +20℃~+30℃ |
Pressure Range | 700hPa~1060hPa |
Service Altitude | ≤2000m |
Over Voltage Category | II |
Pollution Levels | 2 |
Ambient Air Velocity | ≤0.3m/s |
Other Indicators | Environmental Noise In Incubator Baby Compartment | No more than 55dB(A) at stable temperature |
Carbon Dioxide Concentration In The Infant Compartment | A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization |
The velocity of air flow over a baby's mattress | ≤0.35m/s |
General Alarm | Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm |
The System Alarm | Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure |