Gas Chromatograph BK-GC900A
Brand BIOBASE
Product origin CHINA
Delivery time Within 7 days
Supply capacity Factory Direct Supply
Model BK-GC900A
Introduction:
This instrument, through its meticulously designed and optimized sample introduction system, significantly improves the accuracy and repeatability of sample analysis. Furthermore, its circuitry has undergone a comprehensive upgrade, not only greatly accelerating computational speed and data processing capabilities but also significantly enhancing the instrument's stability and durability, creating a more robust and efficient analytical platform for users. This enables the instrument to maintain long-term stable operation in various application environments, easily meeting users' diverse daily testing needs.
Download
Features:
1. Featuring a 7-inch high-definition capacitive touchscreen, it can display spectra, all temperature zones, and detector functions in real time.
2. Main module design: Can simultaneously install three injection ports and four detectors.
3. Pre- and post-chromatographic channels can be controlled and operated synchronously or asynchronously.
4. Certified with rigorous functional safety ECM, comprehensively enhancing anti-interference capabilities.
Technical Parameters:
Model | BK-GC900A | |
Oven | Temperature Range | RT+5~450℃ |
Temperature Control Accuracy | 0.01℃ | |
Temperature Program | ≥20 levels(expandable) | |
Maximum Temperature Heating Rate | 120℃/min | |
Temperature Control System | An 8-channel independent temperature control system with automatic temperature protection function | |
Selectable Detector Types | Flame Ionization Detector (FID) | Operating temperature: 450℃ |
Detection limit ≤ 5.0*10-12 g/s (n-hexadecane/isooctane) | ||
Baseline drift (30 min) ≤ 2*10⁻¹³ A | ||
Baseline noise ≤2*10⁻14A | ||
Dynamic linear range ≥10⁷ | ||
Automatic ignition function | ||
Thermal Conductivity Detector (TCD) | Sensitivity≥8,000 mV·mL/mg | |
Baseline drift (30min): ≤0.2mV | ||
Baseline noise: ≤0.05mV | ||
Dynamic linear range: ≥10⁵ | ||
Safety protection: Gas bridging current protection function | ||
Selectable Sample Injection Method | It can be used with standard and inert flow path split/splitless capillary injection, septum purge packed column injection, cold column injection, programmable temperature vaporization injection ports, valve injection systems, headspace injection systems, thermal desorption injection systems, etc. | |
Standard Accessories | FID detector, chromatography workstation software, capillary sampler | |
Optional Accessories | Sample injection system, TCD detector, chromatographic column, nickel catalytic converter | |
Power Supply | AC220V, 50/60Hz | |
550*550*480mm | ||
Net Weight | 45kg | |
Packing Size(W*D*H) | 750*710*700mm | |
Gross Weight | 57kg | |
Remark: 2 different types of detectors can be installed simultaneously.