Auto Hematology Analyzer BK-3200
  • Auto Hematology Analyzer BK-3200
  • Auto Hematology Analyzer BK-3200
  • Auto Hematology Analyzer BK-3200

Auto Hematology Analyzer BK-3200

Brand BIOBASE

Product origin CHINA

Delivery time Within 7 days

Supply capacity Factory Direct Supply

Features
1. 9.7-inch color touch LCD screen, easy to operate, display all parameters and histograms on the same screen
2. Fewer samples are required for testing: whole blood≤12μl, pre-diluted≤20μl
3. 60 samples/hour.
4. 21 parameters + 3 histograms
5. It can be turned on continuously for 24 hours
6. Automatically clean the sampler, counting pool and pipeline after starting up
7. Built-in thermal printer, external printer can be connected.
8. Electrical impedance method, cyanide-free detection of HGB
9. Fully intelligent blockage removal and alarm function
10. Both automatic calibration and manual calibration are available
11. Counting Mode: Venous, capilary, prediluted
12. Support LIS system

Download

Auto Hematology Analyzer BK-3200

Features:

1. 9.7-inch color touch LCD screen, easy to operate, display all parameters and histograms on the same screen

2. Fewer samples are required for testing: whole blood≤10ul, pre-diluted≤20ul

3. 60 samples/hour.

4. 21 parameters + 3 histograms

5. It can be turned on continuously for 24 hours

6. Automatically clean the sampler, counting pool and pipeline after starting up

7. Built-in thermal printer, external printer can be connected.

8. Electrical impedance method, cyanide-free detection of HGB

9. Fully intelligent blockage removal and alarm function

10. Both automatic calibration and manual calibration are available

11. Counting Mode: Venous, capilary, prediluted

12. Support LIS system


Technical Parameter:

Model

BK-3200

Throughput

60 samples/hour

Principle

Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB

Channels

2

Parameters

21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDWPCT, P-LCR, P-LCC)

Histograms

3 histograms (including WBC, RBC, PLT histograms)

Sample Volume

Whole blood: 12μl, Prediluted: 20μl

Calibration

Manual, auto calibration

Counting Modes

Venous, capilary, prediluted

Printing Model

Auto print, manual print

LIS

Support Bi-directional LIS

Language

Chinese, English, Spanish, French, Russian, Ukrainian

Reagent

Diluent 10L, Hemolytic agent 250ml, Probe cleanser: 50ml

Data Input

9.7-inch touch screen, mouse, keyboard (optional)

Output

Internal printer, support external printer

Printer paper

57*30mm

Interface

4 USB ports, network portRS232 serial port.

Storage

100,000 results with histogram

Unclog

High voltage burn, reverse high-pressure flush

Work Temperature

15-30°C

Power Supply

100~240 AC, 50/60Hz, 150VA

Relative Humidity

30﹪~85

Atmospheric Pressure

70.0kPa-110.0kPa

Carry-over rate

Parameter

WBC

RBC

HGB

PLT

CV

≤ 1.5%

≤ 1%

≤ 1%

≤ 3%

Background

Parameter

WBC

RBC

HGB

HCT

PLT

Background

≤ 0.3×109/L

≤ 0.03×1012/L

≤ 1 g/L

≤ 0.5%

≤ 10×109/L

Linearity

Parameter

Measurement range

CV

WBC

(0.5~10.00×109)/L
(10.10~99.9×109)/L

(±0.5×109)/L
±5%

RBC

(0.05~1.00×1012)/L (1.01~9.99×1012)/L

(±0.05×1012)/L
±5%

HGB

(2~70g)/L
(71~300.0g)/L

(±2.0g)/L
±2%

PLT

(10~100×109)/L
(101~999×109)/L

(±10.0×109)/L
±8%

Precision

Parameter

Whole blood (CV)

Pre-diluted CV 

Measurement range

WBC

≤ 2.0%

≤ 4.0%

(4.0-15.0×109)/L

RBC

≤ 1.6%

≤ 3.0%

(3.5-6.0×1012)/L

HGB

≤ 1.6%

≤ 3.0%

(110.0-180.0g)/L

MCV

≤1.0%

≤2.0%

(70.0-120.0)fL

PLT

≤ 4.0%

≤8.0%

(150.0-500.0×109)/L

External Size(W*D*H)

295*467*418mm

Packed Size(W*D*H)

Instrument: 404*564*616mm; Reagent: 480*480*480mm

Net weight

19 KG

Gross weight

Instrument: 31KG; Reagent::25Kg


Get the latest price? We'll respond as soon as possible(within 12 hours)

Privacy policy

close left right