Automatic Chemistry Analyzer BK-2000
Brand BIOBASE
Product origin CHINA
Delivery time Within 7 days
Supply capacity Factory Direct Supply
Automatic Chemistry Analyzer, a state-of-the-art solution for modern laboratories. With a capacity of 300 sample positions and 216 reaction cuvettes, coupled with 170 refrigerated reagent positions, this analyzer offers unparalleled efficiency and precision.
Download
Features:
1. 300 Sample positions.
2. 432 Reaction cuvettes.
3. 250 Refrigerated reagent positions.
4. Double focus, double aperture, holographic concave grating.
5.2000 Tests per hour, ISE is optional.
6. Built-in sample barcode system.
7. Probe with anti-collision function, liquid level detection function.
Model
Model | BK-2000 | |
Overall | Throughput | 2000Tests/hour, ISE(Optional) |
Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |
Certificates | CE, FDA, ISO9001, ISO14001, ISO13485 | |
Sample & Reagent System | Sample Positions | 300sample positions |
Reagent Position | 250 reagent positions | |
Sample Volume | 1-35μL | |
Reagent Volume | 1-350μL | |
Probe | Teflon coating, with anti-collision function, Liquid level detection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with independent switch | |
Reaction System | Temperature Control | Water bath37±0.1℃, real-time monitoring |
Cuvettes | 432 reusable cuvettes, optical length 5.5mm | |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT function | YES | |
Optical System | Light Source | 12V/50W halogen lamp/ holographic concave grating |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 16 wavelength from 340~850nm | |
Absorbance | 0~3.5Abs | |
Calibration & QC | Calibration | Linear: K factor, 1-point ,2-point and multipoint linear Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P |
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10 , 32or 64bit |
LIS System | Bi-direction, support HL7 protocol | |
Interface | LAN port access | |
Printer | External, multiple reporting mode available | |
Working Conditions | Power Supply | AC220V±10% 50/60Hz(Standard), 110V±10% 60Hz(Optional), 3000W |
Temperature | 15~30℃(±2℃) | |
Water Consumption | Deionized water: 80L/H | |
Humidity | 40~85% | |
Size & Weight | External Size (W*D*H) | Scheduling module: 972*1225*1495mm; Analysis module:1505*1225*1245mm |
Net Weight | Scheduling module: 160kg; Analysis module:500kg | |
Package Size (W*D*H) | Scheduling module:1410*1350*1170mm; Analysis module:1685*1410*1460mm | |
Gross Weight | Scheduling module: 235kg; Analysis module:600kg |